Features
- Improving Portability, Minimizing Structure of Plate Handle and Controller
- Real Time Monitoring and Data Logging in Computer by Free Software
Item Code | M80E-017 | |
Model | ARC-P102ZB | |
Controller | Dimension | 63 mm(H)×125 mm(W)×154 mm(D) |
Weight | < 1.2 Kg | |
Plate | Dimension | 150 mm(H)×150 mm(W)×43 mm(D) |
Capacitance | 20pF ± 2pF | |
Weight | < 0.8 Kg | |
Sensor | Oscillating chopper type (DIT) | |
Input Voltage | 12 Vdc ±10% | |
Input Current | 1A | |
Battery Specification | NiMH 7.2V 6 pack | |
Battery Charging Time | 3 hrs (Until full charge) | |
Battery Use Time | 8 hrs (Full charge statement) | |
Float Mode | Range | 0 to ±1100V DC or peack AC |
Bandwidth (-3dB) | DC to 80 Hz | |
Stability | < 10% of initial Voltage with 5 minutes [40% RH at 25℃] |
|
Decay-time Mode |
Start/Stop Accuracy | 2 L/min (±10%) [Under 0.1MPa] |
Time Resolution | 0.1 sec (0.1~999.9 sec) | |
Range | Start Voltage: ±1~±1000V / Available 1V intervers. Stop Voltage: 0~±999V / Available 1V intervers. |
|
Material | Controller | SUS 1.2 T |
Plate | SUS 1.2 T | |
Function | (+) Decay-time, (-) Decay-time, Reset | |
Display | FND 2 set (Plate-voltage, Decay-time) | |
Communication | USB communication | |
Ambient Temperature | 5℃ ~ 35℃ | |
Ambient Humidity | RH 80% under (No condensation) | |
Storage Temperature | 0℃~ +50℃ (32 ~ 122℉) | |
Storage Humidity | RH 80% under (No condensation) |